Multi-Gas Monitor
Model | MD-811 | ||
---|---|---|---|
Measuring Gas | Oxygen | Combustible Gases | Carbon monoxide |
Measuring Principle | Galvanic cell | Catalytic combustion | Chronoamperometry |
Measuring Range | 0 - 50.0 vol% | 0 - 100%LEL | 0 - 300ppm |
Resolution | 0.1 vol% | 1%LEL | 1ppm |
Accuracy | Within ±0.5 vol% at below 25.0 vol%/Within ±3.0 vol% at over 25.0 vol% | Within ±10%LEL at below 100%LEL | Within ±10ppm at 0 - 100ppm/±10% of indicated value at over 101ppm |
Display Method | Digital liquid crystal | ||
Alarm Setting | At below 18% | At over 20%LEL | At over 50ppm |
Alarm Method | LED lamp, LED display and buzzer | ||
Response Time | 90% of the response within 25 seconds (excluding gas inlet tube) | ||
Power Source | 4 size D alkaline batteries/AC 100V (adaptor) | ||
Continuous Operation | Up to 35 hours with alkaline batteries | ||
Operating Temperature Range | Temperature : -10 - 40℃, Humidity : 30-80%RH (non condensing) | ||
Dimensions | 230(W)×165(H)×130(D)mm | ||
Weight | 2.8kg (including dry batteries and sensors) | ||
Standard Accessories | Sampling tube (8m) with float-type sampler, leather case |